MEDICAL FREEZERS
Models
AFZ2PVBIADA
AFZ5PVBIADA
AFZ7PV
AFZ14PV
AFZ17PV
AFZ19PV
User Manual
BEFORE USE, PLEASE READ AND FOLLOW ALL SAFETY RULES AND OPERATING
INSTRUCTIONS
Accucold Division of Felix Storch, Inc.
An ISO 9001:2015 registered company
770 Garrison Avenue
Bronx, NY 10474
www.accucold.com

2
EQUIPMENT SAFETY
Your safety and the safety of others are very important.
We have provided many important safety messages in this manual and on your unit. Always read and
obey all safety messages.
This is the Safety Alert Symbol. The symbol alerts you to potential hazards that
can kill or injure you and others. All safety messages will follow the Safety
Alert Symbol and either the words "DANGER" or "WARNING".
DANGER means that failure to heed this safety
statement may result in severe personal injury or
death.
WARNING means that failure to heed this safety
statement may result in extensive product
damage, serious personal injury, or death.
WARNING: Low temperature/freezing conditions, frostbite HAZARD
WARNING: flammable material
All safety messages will alert you about the potential hazard, tell you how to reduce the chance of injury,
and let you know what can happen if the instructions are not followed.
EQUIPMENT SAFETY
Before the equipment is used, it must be properly positioned and installed as
described in this manual, so read the manual carefully. To reduce the risk of fire,
electrical shock or injury when using this equipment, follow basic precautions,
including the following:
•
Plug into a grounded 3-prong outlet, do not remove grounding prong, do not use an adapter, and do
not use an extension cord.
•
Replace all panels before operating.
•
It is recommended that a separate circuit serving only your unit be provided. Use receptacles that
cannot be turned off by a switch or pull chain.
•
Never clean the equipment parts with flammable fluids. These fumes can create a fire hazard or
explosion. And do not store or use gasoline or other flammable vapors and liquids in the vicinity of
this or any other pieces of equipment. The fumes can create a fire hazard or explosion.
•
Before proceeding with cleaning and maintenance operations, make sure the power line of the unit is

3
disconnected.
•
Do not connect or disconnect the electric plug when your hands are wet.
•
Unplug the unit or disconnect power before cleaning or servicing. Failure to do so can result in
electrical shock or death.
•
Do not attempt to repair or replace any part of your unit unless it is specifically recommended in this
manual. All other servicing should be referred to a qualified technician.
•
These units are CFC- and HFC-free and contain a small quantity of propane (R290). They are
environmentally friendly, but flammable. They do not damage the ozone layer, nor do they increase
the greenhouse effect. Care must be taken during transportation and setting up of a unit that no parts
of the cooling system are damaged. Leaking coolant can ignite and may damage the eyes.
•
In the event of any damage:
-
Avoid open flames and anything that creates a spark,
-
Disconnect from the electrical power line,
-
Air the room in which the unit isolated for several minutes, and
-
Contact the Service Department for advice.
•
The more coolant there is in a unit, the larger the room it should be installed in. In the event of a
leakage, if the unit is in a small room, there is the danger of combustible gases building up. For every
ounce of coolant, at least 325 cubic feet of room space is required. The amount of coolant in the unit
is stated on the data plate inside the unit. It is hazardous for anyone other than an Authorized Service
Person to carry out servicing or repairs to this piece of equipment.
•
Take serious care when handling, moving, and using the unit to avoid either damaging the refrigerant
tubing or increasing the risk of a leak.
•
Replacing component parts and servicing shall be done by factory-authorized service personnel so as
to minimize the risk of possible ignition due to incorrect parts or improper service.
FOLLOW WARNING CALLOUTS BELOW ONLY WHEN APPLICABLE TO YOUR MODEL
•
Use two or more people to move and install unit. Failure to do so can result in back or other injury.
•
To ensure proper ventilation for your unit, the front of the unit must be completely unobstructed.
Choose a well-ventilated area with temperatures above 60°F (16°C) and below 90°F (32°C). [For
optimal performance, install the unit where the ambient temperature is between72º and 78ºF (23º-
26ºC).] This unit must be installed in an area protected from the elements, such as wind, rain, water
spray or drips.
•
The unit should not be located next to ovens, grills or other sources of high heat.
•
The unit must be installed with all electrical, water and drain connections in accordance with state and
local codes. A standard electrical supply (110 V AC only, 60 Hz), properly grounded in accordance
with the National Electrical Code and local codes and ordinances, is required.
•
Do not kink or pinch the power supply cord of the unit.
•
The size of the fuse (or circuit breaker) should be 15 amperes.
•
It is important that the equipment be leveled in order to work properly. You may need to make several
adjustments to level it.
•
All installations must be in accordance with local plumbing code requirements.
•
Make certain that the pipes are not pinched, kinked or damaged during installation.
•
Check for leaks after connection.
•
Never allow children to operate, play with or crawl inside the unit. Medical device, only professional use.
•
Do not use solvent-based cleaning agents or abrasives on the interior. These cleaners may damage
or discolor the interior.
•
Use this equipment only for its intended purpose as described in this Instruction Manual.
•
Keep fingers out of the “pinch point” areas. Clearances between the door and cabinet are necessarily
small. Be careful closing the door when children are in the area.
•
Ensure all ventilation openings are not obstructed.
•
Do not use mechanical devices or other to accelerate the defrosting process, other than those
recommended by the manufacturer.
•
Do not damage the refrigerant circuit.
Termékspecifikációk
Márka: | Accucold |
Kategória: | mélyhűtő |
Modell: | AFZ1PVLHD |
Szüksége van segítségre?
Ha segítségre van szüksége Accucold AFZ1PVLHD, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek
Útmutatók mélyhűtő Accucold
29 Március 2025
25 December 2024
25 December 2024
25 December 2024
25 December 2024
29 Szeptember 2024
29 Szeptember 2024
29 Szeptember 2024
29 Szeptember 2024
29 Szeptember 2024
Útmutatók mélyhűtő
- mélyhűtő NewAir
- mélyhűtő Haier
- mélyhűtő OK
- mélyhűtő Icecool
- mélyhűtő Summit
- mélyhűtő LG
- mélyhűtő Samsung
- mélyhűtő Frigidaire
- mélyhűtő Husqvarna
- mélyhűtő Whynter
- mélyhűtő Cookology
- mélyhűtő Thermador
- mélyhűtő Edgestar
- mélyhűtő Philco
- mélyhűtő Insignia
Legújabb útmutatók mélyhűtő
10 Április 2025
10 Április 2025
10 Április 2025
9 Április 2025
9 Április 2025
6 Április 2025
30 Március 2025
22 Március 2025
19 Március 2025
19 Március 2025