Instructions to User 
Dear Users, thank you very much for purchasing our product. 
This Manual is written and compiled in accordance with the
 council directive IEC 60601-1、IEC 60601-1-11 and 
ISO 80601-2-61 for medical devices and harmonized standards. In case of modifications and software upgrades, 
the information contained in this document is subject to change without notice. 
The  Manual  describes,  in  accordance  with  the  Pulse  Oximeter’s  features  and  requirements,  main  structure, 
functions, specifications, correct methods for transportation, installation, usage, operation, repair, maintenance and 
storage,  etc.  as  well  as  the  safety  procedures  to  protect  both  the  user  and  equipment.  Refer  to  the  respective 
chapters for details. 
Please  read  the  Manual  very  carefully  before  using  this  device.  These  instructions  describe  the  operating 
procedures to be followed strictly, failure to follow these instructions can cause measuring abnormality, equipment 
damage and personal injury. The manufacturer is NOT responsible for the safety, reliability and performance issues 
and any monitoring abnormality, personal injury and equipment damage due to user’s negligence of the operation 
instructions. The manufacturer’s warranty service does not cover such faults.   
Owing to the forthcoming renovation, the specific products you received may not be totally in accordance with the 
description of this User Manual. We would sincerely regret for that.   
The device is a medical device, and it can be used repeatedly.   
Notice : Please read the Manual very carefully before using this device. 
1 Safety 
1.1 Instructions for safe operations 
1) Inspect periodically, make sure that there is no visible damage that may affect patient’s safety and monitoring 
performance about cables and transducers. It is recommended that the device should be inspected at least once a 
week. Please stop using the device if there is obvious damage to the device 
2) Necessary  maintenance  must  be  performed by qualified  service  engineers appointed by  our  company ONLY. 
Users are not permitted to maintain the device by themselves. 
3) The oximeter cannot be used together with devices not specified in User’s Manual. Only the accessory that is 
appointed or recommendatory by manufacture can be used with this device. 
4) The device has been calibrated before leaving factory. 
1.2 Attentions 
 Keep the oximeter away from dust, vibration, corrosive substances, explosive materials, high temperature and 
moisture. 
 If the device gets wet, please stop operating it. 
 When  the  device  is  carried  from  cold  environment  to  warm  or  humid  environment,  please  do  not  use  it 
immediately. 
 DO NOT operate switch on front panel with sharp materials. 
 High temperature or high pressure steam disinfection of the oximeter is not permitted. Refer to User Manual in 
the relative chapter (6.1) for instructions of cleaning and disinfection. 
 Do  not  have  the oximeter immerged in  liquid.  When  it  needs  cleaning,  please wipe  its  surface with medical 
alcohol. Do not spray any liquid on the device directly. 
 When cleaning the device with water, the temperature should be lower than 60 ℃. 
 The fingers which are too thin or too cold may affect the measure accuracy, please insert the thicker finger such 
as thumb or middle finger deeply enough into the probe. 
 Do not use the device on infant or neonatal patients. 
 The product is suitable for children above four years old and adults(Weight should be between 15 Kg to 110 
Kg).   
 The device may not work for all patients. If you are unable to achieve stable readings, discontinue use. 
 The update period of data is less than 5 seconds, which is changeable according to different pulse rate values. 
 If  some  abnormal  conditions  appear  on  the  screen  during  test  process,  pull  out  the  finger  and    reinsert  to 
restore normal use. 
 The device has useful life for three years. 
 The lanyard is made of insensitive material, please don't use it if any person is extra sensitive to lanyard. And 
do not wrap the lanyard around neck to avoid an accident. 
 The device hasn't low-voltage alarm function, it only shows the low-voltage, please change the battery when the 
battery voltage is used up. 
 The device hasn’t the alarm function of exceeding limits, do not use it in situations where alarms are required 
 The  maximum  temperature  for  the  contact  surface  of  the  device  with  the  body  is  less  than  41  ℃,  and  the 
temperature is measured by a temperature measuring device. 
 Batteries must be removed if the device is going to be stored for more than one month, or else batteries may 
leak. 
 Do not twist or pull on the connection circuit. 
1.3 Warnings 
Warning : 
1) Explosive hazard—DO NOT use the device in environment with inflammable gas such as some ignitable 
anesthetic agents. 
2) DO NOT use the device while the testee is being scanned by MRI or CT. 
3) In optical treatment, please follow doctor requirement to use the device. 
4) DO  NOT  strand  the  lanyard  in  order  to  avoid  device  drop  and  damage.  The  lanyard  is  made  of  insensitive 
material. Please do not use it if any person is allergic to lanyard. Do not wrap the lanyard around neck to avoid 
an accident. 
5) The person who is allergic to rubber can not use this device. 
6) The disposal of scrap device and its accessories and packing(including battery, plastic bags, foams and paper 
boxes) should follow the local laws and regulations. 
7) Please check the packing before use to make sure the device and accessories are totally in accordance with the 
packing list, or else the device may have the possibility of working abnormally.   
8) Please don't measure this device with function tester for the device's related information. 
9) The uncomfortable or painful feeling may appear if using the device ceaselessly, especially for the 
microcirculation barrier patients. It is recommended that the sensor should not be applied to the same finger for 
over 2 hours. 
10) The SpO2 probe can not be clipped on the edema and tender tissue. 
11) The infrared is harmful to eyes, so the user and the maintenance man should not stare at the light part of the 
SpO2 probe (the infrared is invisible). 
12) Testee can not use enamel or other makeup. 
13) The fingernails of the testee should not be too long. 
14) For  the  details  of  correlative  clinic  restriction  and  contraindications,  please  refer  to  the  related  medical 
literatures. 
15) The device is not intended for treatment. 
16) The device is a medical device, and it can be used repeatedly. 
1.4 Attention for operation 
a. Please check the device before using, and confirm that it can work normally. 
b. The finger should be placed properly (see the attached illustration of this manual, Figure 1), or else it may cause 
inaccurate measurement. 
c. The SpO2 sensor and photoelectric receiving tube should be arranged in a way with the subject’s arteriole in a 
position there between. 
d. The  SpO2  sensor  should  not  be  used  at  a  location  or  limb  tied  with  arterial  canal  or  blood  pressure  cuff  or 
receiving intravenous injection. 
e. Make sure the optical path is free from any optical obstacles like rubberized fabric. 
f. Excessive  ambient  light  may  affect  the  measurement  result.  It  includes  fluorescent  lamp,  dual  ruby  light, 
infrared heater, direct sunlight and etc. 
g. Strenuous action of the subject or extreme electrosurgical interference may also affect the accuracy. 
h. Testee can not use enamel or other makeup. 
i. Please clean and disinfect the device after operating according to the User Manual (6.1.1) 
j. The Pedometer should be worn on the waist, and fixed well. 
1.5 Clinical restrictions 
1. As the measure is taken on the basis of arteriole pulse, substantial pulsating blood flow of subject is required. 
For a subject with weak pulse due to shock, low ambient/body temperature, major bleeding, or use of vascular 
contracting  drug,  the  SpO
2  waveform  (PLETH)  will  decrease.  In  this  case,  the  measurement  will  be  more 
sensitive to interference. 
2. For those with a substantial  amount of  staining dilution drug (such  as methylene blue, indigo  green and  acid 
indigo blue), or  carbon monoxide hemoglobin  (COHb),  or methionine (Me+Hb) or  thiosalicylic hemoglobin, 
and some with icterus problem, the SpO
2 determination by this monitor may be inaccurate. 
3. The drugs like dopamine, procaine, prilocaine, lidocaine and butacaine may also be a major factor blamed for 
serious error of SpO2 measure. 
4. As the SpO2 value serves as a reference value for judgement of anemic anoxia and toxic anoxia, some patients 
with serious anemia may also report good SpO
2 measurement. 
  Notice   
The product is not suitable for use in continuous supervision for patients. 
The  problem  of  overrating  would  emerge  when  the  patient  is  suffering  from  toxicosis  which  caused  by 
carbon monoxide, the device is not recommended to be used under this circumstance. 
2 Principle 
2.1 Operation principle for Pulse Oximeter 
Figure 1 Operating principle 
Principle of the Oximeter is as below: An experience formula of data process is established taking use of Lambert 
Beer Law according to Spectrum Absorption Characteristics of Reductive Hemoglobin (Hb) and Oxyhemoglobin 
(HbO2)  in  glow  &  near-infrared  zones.  Operation  principle  of  the  instrument  is:  Photoelectric  Oxyhemoglobin 
Inspection Technology is adopted in accordance with  Capacity Pulse Scanning & Recording Technology, so that 
two  beams  of  different  wavelength  of  lights  can  be  focused  onto  human  nail  tip  through  perspective  clamp 
finger-type  sensor.  Then  measured  signal  can  be  obtained  by  a  photosensitive  element,  information  acquired 
through which will be shown on screen through treatment in electronic circuits and microprocessor. 
2.2 Operation principle for Pedometer 
The Pedometer, basing on counting steps by acceleration sensor, adopts the recognition principle of pace waveform 
and  its  acceleration  and  deceleration  process for  recognizing  the  waveform  produced  by person  walking,  finally 
gets the number of steps. Detailed method: use the acceleration sensor to collect user’s steps(more than three steps), 
via  analyzing  and  calculating,  obtain  the  peak  value  of  step  vibration  waveform  and  the  average  value  of 
acceleration difference value, and set it to the threshold value. Collect the user’s actual step waveform data, if it is 
in the threshold range, then it is considered that user walks one step forward. 
3 Technical specifications 
3.1 Main performance 
1) Display of SpO2 value. 
2) Display of pulse rate and bar graph. 
3) Low-battery indication. 
4) Storage function of SpO
2 and pulse rate value 
5) Sync time function. 
6) Display of steps, calorie and time. 
7) The pedometer can store data 
8) Display direction can be changed automatically. 
9) Height, weight and target calorie can be set by the server. 
10) Extra low-power consumption setting. 
11) The device will automatically enter to blank screen state when there is no operation for one minute. 
12) The  data  stored  can  be  uploaded  to  APP  by  Bluetooth,  APP  will  upload  the  data  to  CLOUD  platform  for 
analyzing. 
3.2 Main Parameters 
1. SpO2 measurement range: 0 %~100 % 
Accuracy: 70 %~100 %: ±2 % 
0 %~69 %: unspecified 
2. PR measurement range: 30 bpm~250 bpm 
Accuracy: ±2 bpm or ±2 %, whichever is greater. 
3. Resolution:   
SpO2: 1 %   
PR: 1 bpm 
4. Measurement Performance in Weak Filling Condition: SpO
2 and pulse rate can be shown   
correctly when pulse-filling ratio is 0.4 %. SpO
2 error is ±4 %, pulse rate error is ±2 bpm or ±2 %, whichever is 
greater. 
5. Resistance to surrounding light: The deviation between the value measured in the condition   
of man-made light or indoor natural light and that of darkroom is less than ±1 %. 
6. Pedometer measurement: 
Measurement range: 0~65535 steps 
Resolution: one step 
7. Memory function: record 1000 groups of SpO2 data, 224-day pedometer data. 
8. Working voltage: DC 3 V 
9. Optical Sensor: red light (wavelength is 660 nm,6.65 mW)    Infrared (wavelength is 880 nm, 6.75 mW) 
3.3 Environment requirements 
Storage Environment 
a)Temperature: -40  ~℃ ℃ +60   
b)Relative humidity: ≤95 % 
c)Atmospheric pressure: 500 hPa~1060 hPa 
Operating Environment 
a)Temperature: 10  ~℃ ℃ 40   
b)Relative humidity: ≤75 % 
c)Atmospheric pressure: 700 hPa~ 1060 hPa   
4 Device introduction 
4.1 Appearance introduction 
  Button
   Screen
  Battery compartment 
Termékspecifikációk
| Márka: | Arya | 
| Kategória: | Mérőberendezések | 
| Modell: | CMS50D-BT | 
Szüksége van segítségre?
Ha segítségre van szüksége Arya CMS50D-BT, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek
Útmutatók Mérőberendezések Arya
                        
                         5 Augusztus 2024
                        
                    
                                                            Útmutatók Mérőberendezések
- Mérőberendezések Weidmüller
- Mérőberendezések Shimano
- Mérőberendezések Klein Tools
- Mérőberendezések Motrona
- Mérőberendezések Siig
- Mérőberendezések Alecto
- Mérőberendezések Flex
- Mérőberendezések SilverCrest
- Mérőberendezések Digitus
- Mérőberendezések Technics
- Mérőberendezések Ermenrich
- Mérőberendezések ZKETECH
- Mérőberendezések Trotec
- Mérőberendezések Helvi
- Mérőberendezések Qubino
Legújabb útmutatók Mérőberendezések
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025
                        
                    
                                                            
                        
                         3 Április 2025