1AC II
VoltAlert
Instruction Sheet
Read First: Safety Information
Warning
To prevent possible electric shock or personal injury:
• Read all safety information before you use the Product.
• Do not rely on the Tester to detect the presence or absence of voltage on a
conductor for safety purposes. A single pole non-contact voltage detector such as
this Tester is NOT suitable to determine if a circuit is hazardous. A volt meter or
2-pole voltage detector is necessary to confirm the absence of hazardous voltage
prior to commencing work
• Hold the Product behind the tactile barrier.
• Do not use the Product around explosive gas, vapor, or in damp or wet
environments.
• Replace all alkaline batteries with fresh batteries of the same manufacturer and type
to prevent battery leakage.
• Remove the alkaline batteries if the Product is not used for an extended period of
time, or if stored in temperatures above 50 °C. If the batteries are not removed,
battery leakage may result.
• Repair the Product before use if the battery leaks. Battery leakage may create a
shock hazard or damage the Product.
• Alkaline batteries contain hazardous chemicals that can cause burns. If exposure to
chemicals occurs, clean with water and get medical aid.
• If the Tester is used in a manner not specified by the manufacturer, protection
provided by the Tester may be affected.
• Do not use if VoltBeat is not flashing.
• Test on a known live source within the rated ac voltage range of the product, both
before and after use to ensure unit is in good working condition.
• When using the Tester, if tip does not glow, voltage could still be present. The Tester
indicates active voltage in the presence of AC electric fields of sufficient strength
emanating from the conductor. If the field strength is low, the Tester may not provide
indication of live voltages. Lack of an indication occurs if the Tester is unable to
sense the presence of electric field which may be influenced by several factors the
including, but not limited to:
• Shielded wire/cables
• Thickness and type of insulation
• Distance from the voltage source
• Fully-isolated users that prevent an effective ground
• Receptacles in recessed sockets/ differences in socket design
• Condition of the Tester and Batteries
• Do not use if the Tester appears damaged or if the Tester is not operating properly.
Specifically examine the probe tip for cracks or breakage before use. If in doubt,
have the Tester serviced.
• Do not apply more than the rated voltage as marked on the Tester.
• Use caution with voltages above 30 V ac as a shock hazard may exist.
• Comply with local and national safety codes. Use personal protective equipment
(approved rubber gloves, face protection, and flame-resistant clothes) to prevent
shock and arc blast injury where hazardous live conductors are exposed.
Safety: IEC 61010-1: Pollution Degree 2; IEC 61010-2-030: CAT IV 1000 V
Electromagnetic Environment: IEC 61326-1: Portable Electromagnetic Environment;
CISPR 11: Group 1, Class B
Electromagnetic Compatibility: Applies to use in Korea only. Class A Equipment (Industrial
Broadcasting & Communication Equipment)
[1]
[1] This product meets requirements for industrial (Class A) electromagnetic wave equipment
and the seller or user should take notice of it. This equipment is intended for use in business
environments and is not to be used in homes.
Voltage Detection Range:
Model
Region
Range
A1/A2
N. America, S. America, Asia
90 V ac to 1000 V ac
E1/E2
Europe
200 V ac to 1000 V ac
L (low voltage) AC
N. America
20 V ac to 90 Vac
C
China
90 V ac to 1000 V ac
FOR USE BY COMPETENT
PERSONS
Anyone using this instrument should be
knowledgeable and trained about the risks
involved with measuring voltage, especially
in an industrial setting, and the importance of
taking safety precautions and of testing the
instrument before and after using it to ensure
that it is in good working condition.
Termékspecifikációk
| Márka: | Fluke |
| Kategória: | nincs kategorizálva |
| Modell: | 1LAC II |
Szüksége van segítségre?
Ha segítségre van szüksége Fluke 1LAC II, tegyen fel kérdést alább, és más felhasználók válaszolnak Önnek
Útmutatók nincs kategorizálva Fluke
9 Október 2024
8 Október 2024
5 Október 2024
2 Október 2024
29 Szeptember 2024
23 Szeptember 2024
22 Szeptember 2024
22 Szeptember 2024
20 Szeptember 2024
20 Szeptember 2024
Útmutatók nincs kategorizálva
- nincs kategorizálva WestBend
- nincs kategorizálva Einhell
- nincs kategorizálva IGET
- nincs kategorizálva Marquant
- nincs kategorizálva Revamp
- nincs kategorizálva Genaray
- nincs kategorizálva Kipor
- nincs kategorizálva EnGenius
- nincs kategorizálva Tractive
- nincs kategorizálva IVT
- nincs kategorizálva Yongnuo
- nincs kategorizálva Heller
- nincs kategorizálva Beringer
- nincs kategorizálva Advanced Network Devices
- nincs kategorizálva Stanley
Legújabb útmutatók nincs kategorizálva
10 Április 2025
10 Április 2025
10 Április 2025
9 Április 2025
9 Április 2025
9 Április 2025
9 Április 2025
9 Április 2025
9 Április 2025
9 Április 2025